Dionex IonPac AS11-HC-4µm Column


Dionex IonPac AS11-HC-4µm Column

简要描述:

戴安公司生产的仪器主要有离子色谱仪、高效液相色谱仪、快速溶剂萃取仪和氨基酸直接分析仪。其中离子色谱仪凭借着雄厚的技术优势和方法已广泛地应用于各个领域,其市场占有率达80%。与离子色谱相关的文章有90%均出自戴安公司生产的离子色谱仪。   
戴安公司的快速溶剂萃取仪15分钟之内仅用15毫升溶剂既可完成对10g样品的萃取,该技术已在环境、药物、食品和聚合物工业等领域得到广泛应用。

型号 美国戴安 P/N: 078032

Thermo Scientific Dionex IonPac AS11-HC-4?m 

Hydroxide-Selective Anion-Exchange Column

Higher Resolution of a Large Number of Inorganic Anions and Organic Acid Anions in Complex Matrices with Smaller Particle Columns

Achieve the optimal combination of chromatographic speed and resolution using the new Thermo Scientific? Dionex?IonPac? AS11-HC-4?m anion-exchange column. This high-capacity, high-efficiency column provides excellent resolution of a large number of organic acids and inorganic anions in complex sample matrices. Compared to the previous Dionex IonPac AS11-HC column, the new anion-exchange column uses smaller resin particles for more efficient separations resulting in more accurate peak integration and more reliable results. It has similar selectivity and increased peak efficiencies compared to the Dionex IonPac AS11-HC and was specifically designed to resolve a large number of inorganic anions and organic acids in a single run using a hydroxide gradient. Applications run on the Dionex IonPac AS11-HC column can be easily transferred to the Dionex IonPac AS11-HC-4?m column with the benefit of increased peak resolution, which provides more consistent results. The Dionex IonPac AS11-HC-4?m column is available in three formats; 4 × 250 mm, 2 × 250 mm and 0.4 × 250 mm. The capillary format provides reduced eluent consumption, which helps reduce operating costs.

  • Increase chromatographic resolution and quality of results
  • Use for organic acids and inorganic anions in complex sample matrices or uncharacterized samples
  • Recommended for monovalent and divalent organic acids
  • Use the Dionex IonPac AS11 column for fast analysis of organic acids and anions in well-characterized samples
  • Use the Dionex IonPac ICE-AS1 or ICE-AS6 columns for organic acids in high-ionic-strength samples
  • High-pressure Thermo Scientific Dionex ICS-5000+ HPIC? or ICS-4000 HPIC system is required to run at standard flow rates.

The high-capacity Dionex IonPac AS11-HC-4?m column allows injection of more concentrated samples without column overloading or peak broadening, and provides improved separation over the Dionex IonPac AS11-HC column for monovalent carboxylic acids, including quinate, lactate, acetate, propionate, formate, and butyrate. Complex sample matrices such as chemical wastewater effluents and fermentation broth solutions containing a variety of inorganic anions and organic acids can be analyzed using the high resolution Dionex IonPac AS11-HC-4?m column.

The smaller resin particles used in the Dionex IonPac AS11-HC-4?m column produce a backpressure over 3000 psi at the standard flow rates. A high-pressure IC system, such as the Dionex ICS-5000+ HPIC or the ICS-4000 Capillary HPIC is required to operate this column under standard conditions.

Thermo Scientific? Dionex? IonPac? AS11-HC-4?m Anion-Exchange Column Specifications
Dimensions Dionex IonPac AS11-HC-4?m Capillary Column
0.4 × 250 mm
Dionex IonPac AG11-HC-4?m Capillary Guard Column
0.4 × 50 mm
Dionex IonPac AS11-HC-4?m Analytical Column
2 × 250 mm, 4 × 250 mm
Dionex IonPac AG11-HC-4?m Guard Column
2 × 50 mm, 4 × 50 mm
Maximum Operating Pressure 5000 psi (34 MPa)
-Phase Compatibility pH 0–14; 0–100% HPLC solvents
Substrate Characteristics Bead Diameter: 4.0 ?m supermacroporous resin
Pore Size: 2000 ?
Cross-Linking (%DVB): 55%
Latex Characteristics Functional Group: Alkanol quaternary ammonium ion
Latex Cross-Linking: 6%
Latex Diameter: 70 nm
Hydrophobicity: Medium Low
Capacity 2.9 ?eq (0.4 × 250 mm capillary column)
0.07 ?eq (0.4 × 50 mm capillary guard column)
72.5 ?eq (2 × 250 mm analytical column)
1.75 ?eq (2 × 50 mm guard column)
290 ?eq (4 × 250 mm analytical column)
7 ?eq (4 × 50 mm guard column)
Column Construction PEEK? with 10-32 threaded ferrule-style end fittings
All components are nonmetallic

Analytical Columns
Dionex IonPac AS11-HC-4?m Capillary Column (0.4 x 250 mm) 078031
Dionex IonPac AS11-HC-4?m Analytical Column (2 x 250 mm) 078035
Dionex IonPac AS11-HC-4?m Analytical Column (4 x 250 mm) 082313
Guard Columns
Dionex IonPac AG11-HC-4?m Capillary Guard Column (0.4 x 50 mm) 078032
Dionex IonPac AG11-HC-4?m Guard Column (2 x 50 mm) 078036
Dionex IonPac AG11-HC-4?m Guard Column (4 x 50 mm) 078034